Compound-Blocked Calibration Transfer in a Three-API Finite-Cluster GIAO-DFT 13C NMR Benchmark
Calibration models can improve agreement between calculated and experimental 13C NMR shifts without necessarily transferring
to a new compound. We reanalyzed a literature-derived
finite-cluster
GIAO-DFT
benchmark comprising carbamazepine
(CBZ), theophylline (THP), and ibuprofen (IBP) using compound-blocked
validation and added an independent protocol-heterogeneous
external challenge using lamivudine Form II. Uncalibrated predictions, a one-parameter
offset, two-parameter
affine scaling, and a three-parameter
quadratic model were compared. Across the 20 primary sites, raw RMSE was 1.311 ppm.
Leave-one-
compound-
out
aggregate RMSE was 1.065 ppm for offset correction, 1.113 ppm for affine scaling, and 1.052 ppm for
quadratic calibration; macro-averaged
RMSE was 0.915, 0.986, and 0.916 ppm, respectively. Offset correction was lower than
affine scaling in all three internal folds. In the locked lamivudine challenge (eight 13C sites), however, raw RMSE was 4.174 ppm,
and all three transferred calibrations worsened error: 4.643 ppm offset, 4.623 ppm affine, and 4.592 ppm quadratic. Thus the
external API does not support generalization of the internally favorable calibration behavior. Because the external calculation
used a published Gaussian09 GIAO workflow rather than the historical finite-cluster
protocol, it is an out-of-
domain
robustness
challenge rather than like-for-
like
validation. Fbias is interpreted only retrospectively. The results support a bounded conclusion:
Calibration rankings in this small benchmark are fragile to API and protocol domain shift and require prospective validation
under a harmonized computational workflow.